SFC Prep 150 Mgm System

SFC Prep 150 Mgm System

A greener approach to high-throughput purification

A greener approach to high-throughput purification

Many of today’s laboratories are looking to adopt a greener approach to high throughput purification. Designed for high-throughput, semi-prep to preparative scale purifications, the SFC Prep 150 Mgm System is made for repeat injection (bulk purification) of chiral and achiral compounds.

The SFC Prep 150 Mgm System operates at flow rates higher than traditional high-performance liquid chromatography (HPLC) systems without significant increases in back pressure – allowing for high-speed supercritical fluid chromatography (SFC) purification, improved resolution of compounds, and higher purity isolations.

SFC Prep 150 Mgm System SFC Prep 150 Mgm System

Specifications

Operating flow rate range

20 to 150 mL/min, combined CO2 and co-solvent

Maximum operating pressure

4350 psi (300 bar)

Recovery

>90% (within total flow rate range)

*Individual component specifications may allow operation outside the total system defined parameters. The system is designed and tested only for the ranges defined in system specifications.

Ambient operating temperature

15 to 40 °C

Ambient operating humidity

20 to 80% RH at 40 °C (104 °F)

Pressure rating

10,000 psi

Max operating pressure

8700 psi

Flow range

20 to 200 g/min

Wetted materials

316 stainless-steel, 440C, 17–4PH stainless-steel, sapphire, GFPM, PTFE, AL

Number of solvents

(Co-solvent pump only)

Four solvents on the Solvent Select Valve (SSV)

Solvent conditioning

(Co-solvent pump only)

Sparging one channel per solvent

Operating flow rate range

0.50 to 150.00 mL/min, in 0.01 mL/min increments

Maximum operating pressure

6000 psi up to 100 mL/min, roll off to 5000 psi at 150 mL/min

Operating flow rate range

0.50 to 150.00 mL/min, in 0.01 mL/min increments

Plunger seal wash

Integral, active, programmable

Delay volume

<6.5 mL

Compressibility compensation

Automatic and continuous

Priming

Self prime from solvent container on floor with dry lines

Compositional accuracy

+/-3% absolute (full scale) from 5% to 95%; and from 1 to 150 mL/min

Compositional precision

Retention time variation <0.15 min SD (uracil/caffeine; 1, 5, 20, 50, 150 mL/min; 70:30 water/methanol dial-a-mix)

Default syringe

5 mL

Default loop size

2 mL (Optional; 5 mL, and 10 mL)

Injection volume range

0.010 mL (10 µL) to 9.5 mL (9500 µL)

Sample delivery precision

Less than 1.0% RSD; 0.75 mL injection volume, 2.0 mL loop

Injection linearity

R2 greater than 0.995 using 20–75% of loop volume (2 mL loop)

Sample carryover

No Ketoprofen detected, therefore carry over is less than 0.01%

Primary wetted materials

316 stainless-steel, PEEK, PTFE, borosilicate glass

Number of fractions

1–6 fractions and 1 waste

Wetted material

316 stainless-steel, PEEK

Total flow range

20–150 mL/min

Co-solvent max flow rate

75 mL/min

Operational pressure range

(Optimal)

0–100 psi

50 psi

Max pressure

100 psi

Wetted materials

316 stainless-steel, PEEK, borosilicate glass

Max pressure

4400 psi

Ambient operating temperature

15 to 40 °C

Ambient operating humidity

20 to 80% RH at 40 °C (104 °F)

Wetted materials

316 stainless-steel, carbon filled PEEK, platinum, Teflon

Max pressure

1600 psi

Ambient operating temperature

4–40 °C

Flow rate range

0.0–40.0 mL/min

Wetted materials

Synthetic ruby, sapphire, UHMWPE, PTFE, stainless-steel

Ambient operating temperature

15 to 40 °C

Ambient operating humidity

20 to 80% RH at 40 °C (104 °F)

Pressure rating

10,000 psi

Wetted materials

316 stainless-steel

Wavelength range

190 to 800 nm

Light source

Prealigned, Intelligent Technology Deuterium lamp

Wavelength accuracy

±1.0 nm

Linearity range

<5% at 2 AU (Propylparaben 257 nm, 10 mm cell)

Optical resolution

1.2 nm

Noise, wet

≤60 μAU (254 nm, 2 Hz, 1 s TC, 3.6 BW res, 10 mm analytical cell)

Drift, dry

≤5000 μAU/h (2 h warm-up, constant temp. and humidity at 230 nm, 3.6 BW res, 2 Hz)

Data rate

Up to 80 Hz

Path length

3 mm (semi-prep flow cell)

Cell volume

18.3 μL (semi-prep flow cell)

Pressure limit

6000 psi

Wetted materials

316 stainless-steel, PEEK, fused silica

Line voltage

100 to 240 VAC

Dimensions: D x W x H

Weight: lbs

Modifier stream injector

14.4" x 12.4" x 11.5"

18.0 lbs

Prep fraction selector

14.7" x 9.2" x 6.6"

14.1 lbs

Gas-liquid separator

6.5" x 6.5" x 14.5"

2.6 lbs

ABPR

17.5" x 6.9" x 6.0"

16.0 lbs

P200X

24" x 14.3" x 12.3"

84.0 lbs

Heat exchanger

20" x 4.4" x 4.6"

12.8 lbs

External control

ChromScope 2.1

External communications

Ethernet and RS-232 communications

Event input/outputs

Rear panel contact closure and/or TTL inputs/outputs

Operating system

Windows 10

Voltage range

Frequency

VA

Fuse

Modifier stream injector

240

50

200

F3A/250 V

Prep fraction selector

240

50

200

F3A/250 V

ABPR

240

50

200

F3A/250 V

P200X

240

50

200

F3A/250 V

Heat exchanger

240

50

1000

F3A/250 V

Make-up pump

240

50

220

F3A/250 V

2998 PDA detector

240

50

185

F3.15A/250 V

Julabo chiller

240

50

425

F3.15A/250 V

2545 QGM pump

240

50

650

F6.5A/250 V

Overview

  • Achieve high-speed SFC purification and improved resolution of compounds, leading to higher purity isolations
  • Reduce normal-phase solvent usage and disposal
  • Benefit from increased usability with the new intuitive graphical interface of Chromscope 2.1 SFC Software supported by Windows 10
  • Achieve faster equilibration, lower pressure drops across the column, solvent reduction, and lower cost per sample with Waters SFC Prep 150 Mgm

Recommended Use: For laboratories seeking to adopt a greener approach to high-throughput purification.


Features Header

Features Header






A greener approach to high-throughput purification

The Waters SFC Prep 150 Mgm System is perfectly suited for laboratories seeking to adopt a greener approach to high throughput purification. The SFC Prep 150 Mgm System is the ideal choice when reduction of normal-phase solvent usage and disposal is a critically important step for purification laboratories wanting to become more environmentally friendly.

SFC systems utilize liquid carbon dioxide (CO2) as its main mobile phase in combination with one or more organic solvents, resulting in faster equilibration, lower pressure drops across the column, solvent reduction, and lower cost per sample. The process is reproducible and applicable to a wide range of compounds relevant in the pharmaceutical, life sciences, chemical materials food, and environmental markets.


Achieve higher purity isolations

The SFC Prep 150 Mgm System has a maximum flow rate, CO2 and co-solvent combined, of up to 150 mL/minute allowing use of 19 mm and 30 mm columns. It also has capacity for six fraction collection positions (two liters each).

The user can take advantage of a new intuitive graphical interface controlled by the Waters ChromScope 2.1 SFC Software supported by Windows 10. The SFC Prep 150 Mgm System operates at flow rates higher than traditional HPLC systems without significant increases in back pressure, allowing for high-speed SFC purification and improved resolution of compounds, leading to higher purity isolations. ChromScope v2.1 is an updated version of ChromScope Software that enables support of the Waters SFC Prep 150 Mgm System with the ACQUITY QDa Mass Detector. Additionally, this update adds support for ELSD and the Isocratic Solvent Manager.


System features that support increased throughput

The SFC Prep 150 Mgm System is designed to improve productivity with a greener approach to high-throughput purification, with features like:

  • QGM Quaternary Gradient Pump:  This pump delivers flowrates up to a maximum of 150 mL/min serves as the co-solvent delivery device for this system.
  • P200X CO2 Pump:  This high-pressure delivery pump has dual stainless-steel heads with a cam driven sapphire piston assembly; self-priming check valves; pressure sensor; pressure gauge; brushless motor, and a rupture disc assembly. The pump is designed to improve performance with feedback from the pressure sensor and mass flowmeter.
  • Modifier Stream Injector: Designed to support customer requests for modified-stream injection, this device injects one sample at a time using a standard 2 mL loop and a 5 mL syringe. This allows for settable air gaps on either side of the sample, which function as a buffer between the solvent and the sample, decreasing dilution.   
  • Prep Collection Cabinet: This collection cabinet comes with a retractable front door allowing for easy insertion and removal of 2 liter fraction bottles. The collection cabinet has an exhaust port in the upper backside of the cabinet to easily fit laboratories around the world, and comes standard with a removable spill tray. 
  • Effortless integration with the ACQUITY QDa Mass Detector: The intuitive design of the ACQUITY QDa Mass Detector provides the additional data you need without the cost and complexity of traditional mass spectrometry systems and software, giving you the analytical confidence that the selectivity of mass detection brings to your analysis.


Resources

Documents

Documents


Support

Support



Related

Increase throughput and flexibility in preparative liquid chromatography with the extensive portfolio of purpose-built Waters purification solutions and systems.

ChromScope Software is for use on Waters supercritical fluid chromatography (SFC) and liquid chromatography (LC) purification systems. It enables fully automated instrument control, SFC methods development, and sample purification.

Protect your Waters LC or LC-MS system investment and ensure optimum productivity and compliance with the range of plans in FlexCHOICE Instrument Service Coverage.

Ranging from analytical and preparative scale columns, solid-phase extraction (SPE) cartridges and 96-well plates, biomolecule-based solutions, and many supplies and accessories.
Back To Top Back To Top