Performance Verification of a Proposed USP Monograph for Sorbitan Monooleate Using a Gel Permeation Chromatography (GPC) Method With Refractive Index (RI) Detection
Abstract
The United States Pharmacopeia (USP) proposed revisions to the monograph for sorbitan monooleate to update the procedure for assay and add a limit for organic impurities test, employing a GPC method (USP-PF 50(2)). In this work, the proposed GPC method was run on an Arc™ HPLC System with a strong solvent compatibility kit and RI detector. The GPC results met the USP system suitability requirements and acceptance criteria for both the assay and organic impurities analysis in sorbitan monoleate.
Benefits
- Excellent performance of a proposed USP monograph for sorbitan monooleate by meeting the requirements for the assay and limit of organic impurities analysis
- Reliable GPC analysis using the Arc HPLC System with a strong solvent compatibility kit and RI detector
Introduction
Sorbitan monooleate, also known as Span 80, is a nonionic surfactant used as a stabilizer and emulsifier in cosmetic, pharmaceutical, and food products.1 It helps to stabilize formulations by creating a stable mixture between immiscible ingredients, such as oil and water. Additionally, sorbitan monooleate increases stability of food products and helps to elevate their texture and consistency.1
The USP is updating monographs for Chemical Medicines and excipients across the compendia with new methodologies and technologies. As part of the modernization efforts, the USP proposed revisions to the monograph for sorbitan monooleate.2 The proposed USP monograph updates an assay procedure with a GPC method designed for analysis of sorbitan tri-/higher esters, sorbitan diesters, and sorbitan monoesters. Additionally, the USP proposes addition of a limit test for organic impurities utilizing the same GPC method conditions as that used for the assay.
In this work, the proposed USP monograph for sorbitan monooleate was run on a GPC system composed of an Arc HPLC System with a strong solvent compatibility kit and RI detector. The experimental work, chromatographic separation, and calculations were performed according to the proposed USP monograph assay and limit of organic impurities procedures. The success of the analysis was measured by comparing GPC results against the USP requirements and acceptance criteria for assay and organic impurities analysis. Empower™ Software was used for data acquisition and analysis.
Experimental
Solutions preparation and experimental conditions proceeded as described in the proposed USP monograph for sorbitan monooleate.2
Materials
Tetrahydrofuran (THF) HPLC grade, no preservatives, purchased from Fisher Chemicals, Catalog No.: T425–4. Isopropyl alcohol (IPA) purchased from Honeywell, catalog number LC323–4. Sorbitan monooleate purchased form Sigma-Aldrich.
Sample Description
Standard Solutions
Standard solution for the assay was prepared by dissolving each of oleic acid, 1,4-sorbitan, and isosorbide in THF at 1.0 mg/mL. For organic impurities, preparation of the standard solution proceeded as described for the assay.
Sample Solutions
Sample solution for the assay testing was prepared by dissolving sorbitan monooleate in THF at 1.0 mg/mL. For organic impurities, preparation of the sample solution proceeded as described for the assay.
Method Conditions
System: |
Arc HPLC System with quaternary solvent manager (QSM), flow through needle (FTN) sample manager, and strong solvent compatibility kit (p/n: 205002572). Column heater/cooler (p/n: 186179100) |
Detector: |
Refractive Index (RI) · Flow cell temperature: 30 °C · Sampling rate: 10 pts/sec · Polarity: positive |
Mobile phase: |
Tetrahydrofuran |
Separation: |
Isocratic |
Columns: |
Columns with 7.8 x 300 mm with 5 µm, connected in series using a joining tube (p/n: WAT084080) supplied with columns. 1. Styragel™ HR 1, 100 Å, molecular weight range: 100–5,000 (p/n: WAT044234) 2. Styragel HR 0.5, 50 Å, molecular weight range: 0–1,000 (p/n: WAT044231) |
Column temperature: |
30 °C |
Sample temperature: |
25 °C |
Flow rate: |
0.9 mL/min |
Injection volume: |
20 µL |
Run time: |
30 minutes |
Vials: |
LCMS Maximum Recovery 2 mL volume (p/n: 600000670CV) |
Wash solvents: |
Sample manager/purge wash: tetrahydrofuran Seal wash: isopropyl alcohol |
Data Management
Chromatography software: |
Empower 3 Feature Release 5 Service Release 5 (FR5 SR3) for data acquisition and analysis. |
Results and Discussion
The proposed USP monograph for sorbitan monooleate describes a revised assay procedure and adds a limit for organic impurities testing, both utilizing a GPC method.2 The procedure for organic impurities employs the same standard solution, sample solution and chromatographic conditions as that for the assay. Waters Styragel HR 0.5 and HR 1 columns are recommended for the GPC separation with THF as a mobile phase.
In this work, the GPC analysis was performed following the updated procedures in proposed USP monograph for sorbitan monooleate. Columns were connected in series with the larger pore size column first to reduce back pressure. Calculations were performed using Empower Software.
Peak Assignment
The USP lists relative retention times (RRT) to aid in peak assignment and identification of components for the assay and organic impurities testing (Table 1). These values were used to identify peaks in the chromatographic separation of standard and sample solutions (Figure 1).
System Suitability
System suitability was measured using the standard and sample solutions as indicated in the proposed USP monograph for sorbitan monooleate.2 The results for six replicate injections of the standards solution, showed excellent relative standard deviations (RSD) for peak areas and retention times of ≤ 0.68% and ≤ 0.01%, respectively (Figure 2). A summary of the USP system suitability requirements and results generated by the GPC method for assay and limit of organic impurities are shown in Table 2. The GPC method run on an Arc HPLC System with strong solvent compatibility kit met all the USP acceptance criteria for resolution and RSD of six replicate injections of standard solution.
Assay: Analysis of Sorbitan Tri-/Higher Esters, Sorbitan Diesters, and Sorbitan Monoesters
The percentage (%) of each sorbitan ester component in the sorbitan monooleate sample was calculated by area normalization as instructed by the USP.2 The peak area of individual peak was divided by the sum of the relevant peak areas and multiplied by 100. The results generated by the GPC method for the sorbitan tri-/higher esters, sorbitan diesters, and sorbitan monoesters met the USP acceptance criteria ranges (Table 3).
Limit of Organic Impurities
The percentage of each impurity peak in the sorbitan monooleate sample was determined by comparing the area of each peak to the sum of the relevant peaks. The GPC results were within the USP limits for organic impurities content (Table 4).
Conclusion
The GPC method described in the proposed USP monograph for sorbitan monooleate was successfully run on the Arc HPLC System with a strong solvent compatibility kit and refractive index detector, meeting all the requirements for system suitability, assay, and limit of organic impurities testing. The GPC method demonstrated excellent relative standard deviations (RSD) for peak areas and retention times of ≤0.68% and ≤0.01%, well below the required acceptance criteria of 5.0%. The USP criteria for resolution were also met.
References
- https://www.cnchemsino.com/blog/sorbitan-monooleate-span-80-properties-and-uses.html.
- United States Pharmacopeia (USP) proposed revision to monograph for sorbitan monoleate, publication PF 50(2). https://doi.usp.org/USPNF/USPNF_M77590_30101_01.html.
720008502, September 2024