For research use only. Not for use in diagnostic procedures.
Phosphorylated and carboxylate lipid species are metal sensitive and can readily adsorb to stainless steel surfaces within the flow path of UPLC systems. This process can lead to poor peak shape, low recovery and reduction in sensitivity. Here we show that the ACQUITY PREMIER CSH C18 Column with ACQUITY PREMIER System can significantly improve sensitivity, peak shape, and recovery of phosphorylated and carboxylate lipids compared to ACQUITY standard column and stainless steel ACQUITY UPLC I-Class System.
The ACQUITY PREMIER technology provides:
Phosphatidic acid (PA) is the simplest membrane phospholipid class and have a key function as an intermediate in phospholipid and glycerolipid metabolism. Due to their intermediate function, PA’s are found in small amounts in biological membranes but are crucial signaling molecules in biochemical and physiological processes in cells.1-2 Lysophosphatidic acids (LPA) are biosynthetic precursors with regulatory functions in the mammalian reproduction system. Phosphatidylserines (PS) are important components of cellular membranes, where they comprise 5–10% of total phospholipids with important roles in the regulation of apoptosis, lipid synthesis, and transport. Lysophosphatidylserines (LPS) are signaling phospholipids with an important role during inflammatory processes.1-2 PA and PS are acidic lipids with the potential to carry two negative charges – both charges on the phosphate group for PA, while one negative charge is on the phosphate and the second on the carboxylate group in the case of PS.3
For the last 50 years, stainless steel has been the most commonly used construction material for High Performance Liquid Chromatography (HPLC) instruments and columns. However, stainless steel hardware can negatively impact the peak shape and recovery of some analytes, due to non-specific adsorption and binding. Lipid classes that show these effects typically contain functional groups such as phosphate and carboxylate groups that can form chelation complexes with iron and other transition metal ions. The comprehensive lipidomic analysis of various biological tissues is a challenging task due to the extreme complexity of individual lipid classes varying in their structure, attached functional groups, polarity, dissociation, and ionization behavior.3 Phosphorylated and carboxylate lipids (lipids with phosphate or carboxylic acid functional groups such as glycerophosphates, glycerophosphoserines, glycerophosphoinositols, glycerophosphoinositol monophosphates, glycerophosphoinositol bisphosphates, glycerophosphoinositol trisphosphate, glycerophosphoglycerols, phosphorylated sphingolipids including ceramide-1-phosphate, and Sphingoid base 1-phosphates) are metal sensitive and readily adsorb to stainless steel surfaces present within the flow path of the chromatographic system. This process can lead to poor peak shape, low recovery, and a reduction in sensitivity.
An alternative approach to mitigate interactions of phosphorylated lipids with metal surfaces is to add chelators to the mobile phase or sample. Volatile chelators such as citric acid, acetylacetone, methylphosphonic acid, and medronic acids have been used for LC-MS analyses. However, the use of chelators can negatively impact chromatographic selectivity and MS sensitivity. To address these issues, we have explored the use of a hybrid organic/inorganic barrier surface applied to the metal substrates in UPLC instruments and columns. A hybrid surface technology based on an ethylene-bridged siloxane polymer has been found to be well-suited for reversed-phase (RP) and hydrophilic interaction chromatography (HILIC). The main goal of this application note is to evaluate and compare the chromatographic performance of ACQUITY PREMIER System and ACQUITY PREMIER CSH C18 Column, which incorporates this hybrid surface technology (HST), relative to standard stainless steel surface ACQUITY UPLC I-Class and CSH C18 for phosphorylated and carboxylate lipids.
LPA (16:0/0:0), PA (16:0/18:1), LPS (18:1/0:0), and PS (16:0/18:1) lipid standards, egg chicken PA and brain porcine PS extracts were obtained from Avanti Polar Lipids. A separate stock solution of 1 mg/mL was prepared for PA (16:0/18:1), LPS (18:1/0:0) and PS (16:0/18:1) in chloroform. LPA (16:0/0:0) was prepared in a concentration of 0.5 mg/mL in chloroform/methanol/water (80/20/2) with gentle heating at 40 °C and sonication to facilitate complete dissolution. A working mixture dilution series of 5, 50, 100, and 250 ng/mL was prepared in IPA/ACN (50/50) for the lipid standards. Egg chicken PA and brain porcine PS extracts were obtained as powder and dissolved in chloroform at a concentration of 1 mg/mL. A working solution of the egg chicken PA and brain porcine PS extracts was prepared in IPA/ACN (50/50).
LC Conditions |
|
---|---|
LC system: |
Standard ACQUITY UPLC I-Class and ACQUITY PREMIER |
Detection: |
SYNAPT XS |
Vials: |
Waters Total Recovery UPLC Vials |
Column(s): |
ACQUITY CSH C18 (2.1 x 100 mm, 1.7 µm) ACQUITY PREMIER CSH C18 (2.1 x 100 mm, 1.7 µm) |
Column temp.: |
55 °C |
Sample temp.: |
10 °C |
Injection volume: |
5 µL |
Flow rate: |
400 µL/min |
Mobile phase A: |
600/390/10 (ACN/Water/1M aqueous ammonium formate) in 0.1% formic acid |
Mobile phase B: |
900/90/10 (IPA/ACN/1M aqueous ammonium formate) in 0.1% formic acid |
MS system: |
SYNAPT XS |
Ionization mode: |
ESI- |
Acquisition range: |
100–1200 Da |
Capillary voltage: |
2.5 kV (negative mode) |
Collision energy: |
Linear ramp (Transfer CE) 25–45 eV |
Cone voltage: |
30 V |
Chromatography software: |
MassLynx v4.2 |
MS software: |
MassLynx v4.2 |
LC-MS analysis of phosphorylated and acidic lipids from biological samples is a critical step, especially for low abundant signaling lipid classes such LPA, PA, LPS, and PS (Figures 1A and 1B). Our previously published method4 enables the accurate analysis of most commonly occurring polar and non-polar lipid classes except for the phosphate and carboxylate lipid classes due to very broad tailing peaks and lower sensitivity under these conditions.3-6 The phosphate group of the phosphatidic acids (PA and LPA) and the carboxylate group of the phosphatidylserines (PS and LPS) bind to metal ions to form chelation complexes that causes peak tailing and smearing during chromatographic analysis. To address the issue, we have developed the use of a hybrid organic/inorganic surface technology in the ACQUITY PREMIER CSH C18 (2.1 x 100 mm, 1.7 µm) Column and ACQUITY PREMIER System. A serial dilution of LPA, PA, LPS, and PS was analyzed using standard column with stainless steel surface ACQUITY UPLC System and ACQUITY PREMIER Column with ACQUITY PREMIER System. Figures 2A, 2B, 3A, and 3B show a side by side comparison of extracted ion chromatograms of LPA (16:0/0:0), PA (16:0/18:1), LPS (18:1/0:0), and PS (16:0/18:1) using standard CSH C18 Column with stainless steel surface ACQUITY UPLC and ACQUITY PREMIER CSH C18 Column with ACQUITY PREMIER System. The ACQUITY PREMIER CSH C18 Column with ACQUITY PREMIER System significantly improved the peak tailing and sensitivity of these compounds. The increase in sensitivity can be clearly seen for LPA (Figure 2-A1 and 2-A2) at a lower concentration level (5 and 50 ng/mL) as evidenced by the noticeably larger peak for the ACQUITY PREMIER System, compared to the standard metal surface system. Similarly Figures 2-B1 and 2-B2 (PA), 3-A1 and 3-A2 (LPS), and 3-B1 and 3-B2 (PS) show improved sensitivity, peak shape, and recovery obtained when using ACQUITY PREMIER CSH C18 Column with ACQUITY PREMIER System compared to the standard column and system.
The developed method was applied for the analysis of egg chicken PA and brain porcine PS extracts (Figures 4 and 5). The PA lipid species at m/z 671.47 (16:1_18:1) and 673.48 (16:0_18:1) provide very broad tailing peaks using the standard CSH C18 Column with stainless steel surface ACQUITY UPLC (Figure 4A) compared to ACQUITY PREMIER CSH C18 Column with ACQUITY PREMIER System (Figure 4B). Figures 4A and 4B bottom panels show extracted ion chromatogram for PA (16:0_18:1) with peak width 4.1 min for standard CSH C18 Column with stainless steel (Figure 4A bottom panel) and 1.2 min for ACQUITY PREMIER CSH C18 Column with ACQUITY PREMIER System (Figure 4B bottom panel). Similarly Figure 5 shows PS brain porcine extract under identical conditions in both standard and ACQUITY PREMIER conditions with extracted ion chromatogram for PS (18:0_18:1) with peak width 1.1 min for standard CSH C18 Column with stainless steel (Figure 5A bottom panel) and 0.3 min for ACQUITY PREMIER CSH C18 Column with ACQUITY PREMIER System (Figure 5B bottom panel).
In contrast to most of the previous methods using standard reverse phase columns,6 the ACQUITY PREMIER Technology increases the lipidomics coverage by simultaneous analysis of phosphate and carboxylate containing lipids in addition to other lipid classes without significant binding and peak tailing, which can increase the coverage, sensitivity, and recovery of these biologically important lipid species.
The PREMIER CSH C18 Column with ACQUITY PREMIER UPLC System is a highly reliable method for the analysis of LPA, PA, LPS, and PS lipid molecular species that significantly improves sensitivity, peak tailing, and recovery of phosphorylated and carboxylate lipids. The developed method can be applied to other phosphorylated and carboxylate lipid classes such as phosphatidylinositol phosphates with mono-, di-, and triphosphates, ceramide-1-phosphate, and sphingoid base 1-phosphates.